Model Number 400-254
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Insufficient Information (4580)
Event Date 03/12/2021
Event Type  Injury  
Manufacturer Narrative
The device will not be returned.If additional information becomes available, it will be provided in a supplemental report.
 
Event Description
It was reported a 7mm poly for the ankle was revised and a 400-141 was implanted.
 
Event Description
It was reported a 7mm poly for the ankle was revised and a 400-141 was implanted.
 
Manufacturer Narrative
The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If any further information is provided, the complaint report will be updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TALAR COMP,SINGLE COATED US VERSION SMALL, RIGHT
Type of Device
PROSTHESIS, ANKLE, UNCEMENTED, NON-CONSTRAINED
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
MDR Report Key11887446
MDR Text Key252763732
Report Number0008031020-2021-00241
Device Sequence Number1
Product Code NTG
UDI-Device Identifier00886385016450
UDI-Public00886385016450
Combination Product (y/n)N
PMA/PMN Number
P050050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 08/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/26/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date03/31/2019
Device Model Number400-254
Device Catalogue Number400254
Device Lot Number1407045
Was Device Available for Evaluation? No
Date Manufacturer Received08/26/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age75 YR
-
-