Related manufacturer report number: 2648612-2021-00055.On (b)(6) 2021, a 19mm sjm regent heart valve was selected for implant.While the device was fitted into the aorta and sutures were placed, the device leaflets became fractured.The device was removed with all fractured leaflet pieces and a new 19mm sjm regent heart valve was selected.While the replacement device was being rotated, the valves leaflet became dislodged.The device and leaflet were removed and successfully replaced with another 19mm sjm regent heart valve to resolve the event.The patient remained hemodynamically stable throughout the procedure however, the procedure extended longer than what is clinically acceptable.The patient is stable.
|