• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO BCT TRIMA ACCEL; TRIMA MACHINE, FINAL ASSY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO BCT TRIMA ACCEL; TRIMA MACHINE, FINAL ASSY Back to Search Results
Model Number 81000
Device Problems Mechanical Problem (1384); Device Slipped (1584); Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/20/2021
Event Type  malfunction  
Manufacturer Narrative
Investigation: a terumo bct service technician checked out the machine at the customer site.It was determined that the tuning of the operating mechanism was required.The mechanism was adjusted and the machine was returned to service.It was confirmed that the iv pole collar was installed on this device.The device serial number history report indicates no further related issues have been reported for this device.One year of service history was reviewed for this device with no other problems identified related to the reported condition.Investigation is in process.A follow-up report will be provided.
 
Event Description
The customer reported that the iv pole on the trima equipment unexpectedly lowers down.No injury was reported for this incident and no patient was connected at the time the iv pole was sliding down, therefore no patient information is reasonably known.
 
Manufacturer Narrative
This report is being filed to provide additional information in e.1, h6 and h10.Root cause: the root cause of the reported incident was the locking mechanism on the iv pole needed adjustment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Type of Device
TRIMA MACHINE, FINAL ASSY
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
MDR Report Key11894838
MDR Text Key256230357
Report Number1722028-2021-00194
Device Sequence Number1
Product Code GKT
UDI-Device Identifier05020583810006
UDI-Public05020583810006
Combination Product (y/n)N
PMA/PMN Number
BK190332
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup,Followup
Report Date 05/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number81000
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received06/22/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-