Model Number 777000 |
Device Problems
Output below Specifications (3004); Output Problem (3005)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/07/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The subject device has not yet been returned for evaluation/investigation.Therefore, the root cause of the reported phenomenon could not be determined at this time.This report will be supplemented accordingly following investigations.
|
|
Event Description
|
It was reported that the device pk3 cut set has low output, below the acceptable limits.The issue occurred during device maintenance.There was no patient involvement, no user injury reported due to the event.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A service order was created for this device to be returned and inspected.The customer called in and stated that the device was working again and they would call back if any more issues came up.The service order was closed as no new issues were reported.A dhr (device history record) review has been performed; no issues (ncrs or deviations) with the manufacturing process have been indicated which might explain the failures observed.Olympus will continue to monitor complaints for this device.
|
|