Model Number 119216M
Device Problem Inaccurate Flow Rate (1249)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/07/2021
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that the foley catheter was not draining urine and hence it was retained in the patient.It was also stated that the catheter was not draining even after the foley bag was burped to put air in, and verified that there were no kinks or dependent loops.It was also stated that the temperature foley leaked after insertion and balloon of the catheter was not fully deflating and was leaving a lip or a ridge.
 
Manufacturer Narrative
The reported event was inconclusive.It was unknown whether the device had met relevant specifications.The product was used for treatment.It was unknown whether the product had caused the reported failure.Visual evaluation of the attached photo sample noted one opened (without original packaging) used temperature sensing silicone foley catheter.Visual inspection of the sample noted balloon mushroom was present upon receipt of the photo.Based on the photo sample it was unknown whether the inaccurate flow rate or urine leakage occurred because the device could not be tested.A potential root cause for this failure mode could be due to tooling misalignment.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Sterile: unless package is opened or damaged.Caution: do not aspirate urine through drainage funnel wall.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Warning: this product should never be connected to the temperature monitor or connected to a cable during an mri procedure.Failure to follow this guideline may result in serious injury to the patient.It is important to closely follow these specific conditions that have been determined to permit the examination to be conducted safely.Any deviation may result in a serious injury to the patient.Valve type: use luer slip syringe.Do not use needle.Recommended inflation capacities 8 fr.And 10 fr.(3cc balloon): use 3.5cc sterile water; 12 fr.(5cc balloon): use 5.5cc sterile water; 14 fr.And larger (5cc balloon): use 10cc sterile water.Do not exceed recommended capacities.Caution: aggressive traction, particularly in the presence of suturing, is not recommended for 100% silicone foley catheters.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with applicable laws and regulations.To deflate catheter balloon: gently insert a syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.If you notice slow or no deflation, re-seat the syringe gently.Allow the balloon to deflate slowly on its own.Do not aspirate or manually accelerate the deflation of the balloon.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Visually inspect the product for any imperfections or surface deterioration prior to use.Caution: as with all temperature probes: in the presence of rf energy sources: local heating, temperature errors, and probe damage may occur.In medical use, unplug the temperature-sensing catheter at the extension cable before activating electrosurgical or other types of direct coupled rf energy sources.Do not stretch catheter.This will cause repositioning of probe.Do not use stylet.This will cause stretching of catheter." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Event Description
It was reported that the foley catheter was not draining the urine and hence it was retained in the patient.It was also stated that the catheter was not draining even after the foley bag was burped to put air in and verified that there were no kinks or dependent loops.It was also stated that the temperature foley leaked after the insertion and the balloon of the catheter was not fully deflating and was leaving a lip or a ridge.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRI-SIL® ALL-SILICONE TEMP SENSING FOLEY CATHETER
Type of Device
TEMP SENSING FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key11909304
MDR Text Key253525920
Report Number1018233-2021-03131
Device Sequence Number1
Product Code EZL
UDI-Device Identifier00801741046308
UDI-Public(01)00801741046308
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K070582
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/31/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2023
Device Model Number119216M
Device Catalogue Number119216M
Device Lot NumberNGEZ0900
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/19/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/11/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/25/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-