Model Number N/A |
Device Problems
Difficult to Insert (1316); Peeled/Delaminated (1454)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/06/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that during a hip procedure, the poly would not seat in the cup.The surgeon noticed that the locking mechanism had started peeling.A new poly was opened and was able to be implanted successfully.There was no reported harm or injury to the patient.Attempts have been made and additional information on the reported event is unavailable at this time.
|
|
Event Description
|
No additional event information to report at this time.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: d4; g3; h2; h3; h4; h6.Complaint sample was returned and evaluated against the reported event.Visual examination of the returned product identified the locking feature of the device has been damaged and a piece has been shaved from the liner.Dhr was reviewed and no discrepancies were found.The root cause is unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|