It was indicated that the device will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.
|
Following a non-boston scientific cryo balloon atrial fibrillation procedure in which a blazer prime htd catheter was used, the patient became hypotensive after leaving the room and an echocardiogram revealed a pericardial effusion.A pericardiocentesis was performed to stabilize the patient.The physician did not allege any particular device as causing the pericardial effusion.No perforation was noted during the procedure.The perforation was not able to be located on echocardiogram.During the procedure, the patient's activated clotting time was 390.No radiofrequency lesions were applied during the procedure, only cryo.There were no patient symptoms during the procedure and the procedure had been deemed successful per the physician.The devices were retained by the hospital.
|