• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES, PR SWAN-GANZ CCOMBO V CCO/SVO2/CEDV THERMODILUTION CATHETER; SWAN-GANZ CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES, PR SWAN-GANZ CCOMBO V CCO/SVO2/CEDV THERMODILUTION CATHETER; SWAN-GANZ CATHETER Back to Search Results
Model Number 777F8
Device Problem Material Rupture (1546)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/19/2021
Event Type  malfunction  
Manufacturer Narrative
One catheter with attached monoject 1.5 cc limited volume syringe, three injection ports and two three-way stopcocks was returned for evaluation.A non-edwards contamination shield was located between 59 cm and 80 cm proximal from the catheter tip.Balloon was found to be ruptured at the central area of balloon latex and the ruptured edges were not able to match up.All through lumens were patent without any leakage or occlusion.No visible damage or inconsistency was observed from the catheter body or returned syringe.Visual examination was performed under microscope at 20x magnification and with the unaided eye.A device history record review was completed and documented that device met all specifications upon distribution.Customer report of balloon inflation issue was confirmed.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint and implement any necessary corrective actions.Invasive procedures involve some patient risks.Although serious complications are relatively uncommon, the physician is advised, before deciding to insert or use the catheter, to consider the potential benefits in relation to the possible complications.The techniques for insertion, methods of using the catheter to obtain patient data information, and the occurrence of complications is well described in the literature.It is common clinical practice to check balloon integrity by inflating it to the recommended volume in order to detect any asymmetry or leakage condition before use of the catheter.It is unknown if user or procedural factors may have contributed to the event.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Event Description
It was reported that the balloon of the swan ganz catheter did not inflate during balloon test before use.Patient demographic information requested but unavailable.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ CCOMBO V CCO/SVO2/CEDV THERMODILUTION CATHETER
Type of Device
SWAN-GANZ CATHETER
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer Contact
jessica atallah
1 edwards way
irvine, CA 92614
9492500249
MDR Report Key11915727
MDR Text Key254069004
Report Number2015691-2021-03262
Device Sequence Number1
Product Code DQE
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K040287
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/01/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2022
Device Model Number777F8
Device Catalogue Number777F8
Device Lot Number63371075
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/06/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/11/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/04/2020
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-