STRYKER ORTHOPAEDICS-MAHWAH CUP IMPACTOR BOLT; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
|
Back to Search Results |
|
Model Number 1440-2011 |
Device Problem
Mechanical Jam (2983)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/10/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
|
|
Event Description
|
The connection between the bolt and the trident cup was so tight.When the bolt was removed after the cup was placed in the body, it turned together with the cup.Eventually it could not be press-fitted, so a cement cup was used instead.
|
|