Catalog Number CRE14S
Device Problems Fracture (1260); Detachment of Device or Device Component (2907); Material Split, Cut or Torn (4008)
Patient Problems Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/14/2021
Event Type  Injury  
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer.The investigation of the reported event is currently underway.The catalog number identified has not been cleared in the us but is similar to the crosser cto recanalization catheter products that are cleared in the us.The pro code and 510 k number for the crosser cto recanalization catheter products are identified.(expiry date: 03/2023).
 
Event Description
It was reported that during a recanalization procedure in the anterior tibial artery, the core wire allegedly had fracture and detachment.It was further reported that the tip of the wire was reported to be inside the patient body.There was no reported patient injury.
 
Manufacturer Narrative
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer.The investigation of the reported event is currently underway.H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the crosser cto recanalization catheter products that are cleared in the us.The pro code and 510 k number for the crosser cto recanalization catheter products are identified in d2 and g4.H10: d4 (expiry date: 03/2023), g3 h11: b5, h1, h6(patient).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that during a recanalization procedure in the anterior tibial artery, the core wire allegedly had fracture and detachment.It was further reported that the tip of the wire was reported to be inside the patient body.The current status of the patient is unknown.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the crosser cto recanalization catheter products that are cleared in the us.The pro code and 510 k number for the crosser cto recanalization catheter products are identified in d2 and g4.H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one 14s crosser cto recanalization catheter was returned.During visual evaluation, no anomalies were noted to the transducer handle or saline hub.The distal marker band was attached outer catheter.The core wire was fractured proximal to the marker band and the distal end was not returned.Moreover, tears were noted throughout the outer catheter.Therefore, the investigation is confirmed for tip detachment and torn material.A definitive root cause for the reported tip detachment and identified torn material issue could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 03/2023), g3, h6 (method).H11: b5, h6 (device, result, conclusion).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that during a recanalization procedure in the anterior tibial artery, the core wire allegedly had detachment.It was further reported that the tip of the wire was remained in patient body.The patient status was unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CROSSER CTO RECANALIZATION CATHETER
Type of Device
RECANALIZATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury 12804
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key11921506
MDR Text Key253787366
Report Number2020394-2021-01155
Device Sequence Number1
Product Code PDU
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K161208
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 11/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCRE14S
Device Lot NumberGFFP1402
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/27/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/14/2021
Initial Date FDA Received06/02/2021
Supplement Dates Manufacturer Received07/16/2021
11/16/2021
Supplement Dates FDA Received08/11/2021
11/17/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/05/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Other;
Patient SexMale
-
-