DORNOCH DUO FLUID CART WITH SMOKE EVACUATOR; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
|
Back to Search Results |
|
Model Number N/A |
Device Problem
Smoking (1585)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/17/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).The investigation has been completed.Review of the most recent repair record determined he unit would not power on; the power inlet failed.The plastic of the power inlet was partially melted with the fuses.Additionally, the power cord had a burn mark near when it plugs into the power inlet.The power inlet was replaced and a replacement power cord was sent to the account and resolved the reported issue.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.Device is used for treatment.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.
|
|
Event Description
|
It was reported that unit was tagged with note that it won't power on.Additionally, upon inspection by tech, there were burn marks found on the power cord near where it was plugged into the wall.No adverse events were reported as a result of this malfunction.
|
|
Search Alerts/Recalls
|
|
|