• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DORNOCH DUO FLUID CART WITH SMOKE EVACUATOR; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DORNOCH DUO FLUID CART WITH SMOKE EVACUATOR; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED Back to Search Results
Model Number N/A
Device Problem Smoking (1585)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/17/2021
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).The investigation has been completed.Review of the most recent repair record determined he unit would not power on; the power inlet failed.The plastic of the power inlet was partially melted with the fuses.Additionally, the power cord had a burn mark near when it plugs into the power inlet.The power inlet was replaced and a replacement power cord was sent to the account and resolved the reported issue.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.Device is used for treatment.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.
 
Event Description
It was reported that unit was tagged with note that it won't power on.Additionally, upon inspection by tech, there were burn marks found on the power cord near where it was plugged into the wall.No adverse events were reported as a result of this malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUO FLUID CART WITH SMOKE EVACUATOR
Type of Device
APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
Manufacturer (Section D)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer (Section G)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key11924254
MDR Text Key253850573
Report Number0001954182-2021-00031
Device Sequence Number1
Product Code JCX
UDI-Device Identifier00889024466005
UDI-Public(01)00889024466005
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 06/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00514010200
Device Lot Number0042975
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/17/2021
Initial Date FDA Received06/02/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/20/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-