• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH T.L.I.F. SG INSTRUMENTS RING CURETTE, STRAIGHT 8MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH T.L.I.F. SG INSTRUMENTS RING CURETTE, STRAIGHT 8MM Back to Search Results
Model Number 202711101
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/07/2021
Event Type  malfunction  
Manufacturer Narrative
Reporter is a j&j sales representative.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on (b)(6) 2021, during spinal fusion surgery the instrument broke off.The surgeon was using the instrument to remove disc and the metal tip of the handle broke.The instrument was promptly removed from the sterile field and no harm was committed.There were no fragments generated and no patient consequences.Procedure was completed successfully without any surgical delay.This report is for one (1) t.L.I.F.Sg instruments ring curette, straight 8mm.This is report 1 of 1 for complaint (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6- the complaint condition was confirmed as the device was broken.No definitive root cause could be determined based on the provided information.The unintended forces might have contributed to the reported complaint condition.No new, unique or different patient harms were identified as a result of this evaluation.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device history lot part # 202711101 , lot # 0709hg, supplier: (b)(4).Batch # 1 qty - release to warehouse date: 27 jul 2009.Batch # 2 qty - release to warehouse date: 30 jul 2009.No ncr's were generated during production.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T.L.I.F. SG INSTRUMENTS RING CURETTE, STRAIGHT 8MM
Type of Device
CURETTE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key11929704
MDR Text Key257411345
Report Number1526439-2021-01093
Device Sequence Number1
Product Code HTF
UDI-Device Identifier10705034184234
UDI-Public(01)10705034184234
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 05/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number202711101
Device Catalogue Number202711101
Device Lot Number0709HG
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/24/2021
Initial Date Manufacturer Received 05/07/2021
Initial Date FDA Received06/03/2021
Supplement Dates Manufacturer Received06/16/2021
Supplement Dates FDA Received07/01/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-