Model Number 3650
Device Problem Program or Algorithm Execution Failure (4036)
Patient Problem Arrhythmia (1721)
Event Date 05/19/2021
Event Type  Injury  
Manufacturer Narrative
The reported event of loss of capture during threshold testing was confirmed.The cause of the event was attributed to the programmer's failure to execute the algorithm to terminate testing.No other anomalies were found.Further information was requested but not received.
 
Event Description
It was reported that during capture threshold testing, the user released the button to terminate the testing as expected, however, the device continued to decrement, resulting in a loss of capture.The patient was noted to be in asystole for a few seconds.The test was manually terminated.The patient was stable.
 
Event Description
It was reported the device was subject to the decrement test field action issued by abbott on 10 march 2022.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN PROGRAMMER
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key11948068
MDR Text Key254558453
Report Number2017865-2021-20976
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date 04/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/07/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3650
Device Catalogue Number3650
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/10/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number2017865-03/10/22¿001¿C
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexFemale
-
-