ST. JUDE MEDICAL, INC.(CRM-SYLMAR) UNIFY ASSURA CRT-D RF HV; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3357-40C
Device Problem Over-Sensing (1438)
Patient Problem Heart Failure/Congestive Heart Failure (4446)
Event Date 05/18/2021
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that the patient presented to the hospital on (b(6) 2021 with congestive heart failure (chf).During routine interrogation of the implantable cardioverter defibrillator (icd), it was noted that there were several non-sustained right ventricular oversensing (nsrvo) alerts.After reviewing the electrograms the physician suggested the nsrvo alerts were due to noise.Programming changes were made to the icd.The patient did not suffer any adverse consequences from the nsrvo alert and chf was attributed to non-compliance with a low sodium diet.The patient was in stable condition throughout.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY ASSURA CRT-D RF HV
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key11950857
MDR Text Key254805450
Report Number2017865-2021-21011
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734508186
UDI-Public05414734508186
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 06/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2018
Device Model NumberCD3357-40C
Device Catalogue NumberCD3357-40C
Device Lot NumberA000032146
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/18/2021
Initial Date FDA Received06/07/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/18/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age72 YR
-
-