• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO BCT TRIMA ACCEL; TRIMA ACCEL PLATELET, PLASMA, RBC SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO BCT TRIMA ACCEL; TRIMA ACCEL PLATELET, PLASMA, RBC SET Back to Search Results
Model Number 82406
Device Problems High Readings (2459); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/08/2021
Event Type  malfunction  
Manufacturer Narrative
Investigation: the run data file (rdf) was analyzed for this event.The analysis of the run data file did not find a conclusive cause for the higher than expected wbc content in the platelet product reported for this collection.No unusual process variable was identified and the signals in the run data file indicate that the trima accel system operated as intended.Based on the available information, it cannot be ruled out that the higher than expected wbc content in the platelet product could be donor related.The device history records (dhr) were reviewed for this lot.There were no events noted in the dhr that would have contributed to the elevated wbc count experienced by the customer.Investigation is in process.A follow-up report will be provided.
 
Event Description
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product.No patient (donor) was connected at the time of the event, therefore, no patient information is available.Donor unit # (b)(6).The platelet collection set is not available for return because it was discarded by the customer.
 
Event Description
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the platelet product.No patient (donor) was connected at the time of the event, therefore, no patient information is available.Donor unit#: (b)(4) the platelet collection set is not available for return because it was discarded by the customer.
 
Manufacturer Narrative
Investigation: the run data file (rdf) was analyzed for this event.The analysis of the run data file did not find a conclusive cause for the higher-than-expected wbc content in the platelet product reported for this collection.No unusual process variable was identified and the signals in the run data file indicate that the trima accel system operated as intended.Based on the available information, it cannot be ruled out that the higher-than-expected wbc content in the platelet product could be donor-related.The device history records (dhr) were reviewed for this lot.There were no events noted in the dhr that would have contributed to the elevated wbc count experienced by the customer.A disposable complaint history search was performed for this lot and found 1 report for similar issues on this lot.Root cause: a root cause assessment was performed for this complaint.The analysis of the run data file did not find a conclusive cause for the higher-than-expected wbc content in the platelet product reported for this collection.No unusual process variable was identified and the signals in the run data file indicate that the trima accel system operated as intended.Based on the available information, it cannot be ruled out that the higher-than-expected wbc content in the platelet product could be donor-related.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIMA ACCEL
Type of Device
TRIMA ACCEL PLATELET, PLASMA, RBC SET
Manufacturer (Section D)
TERUMO BCT
lakewood CO 80215
MDR Report Key11959428
MDR Text Key261940755
Report Number1722028-2021-00191
Device Sequence Number1
Product Code GKT
UDI-Device Identifier05020583824065
UDI-Public05020583824065
Combination Product (y/n)N
PMA/PMN Number
BK190332
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 05/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/01/2023
Device Model Number82406
Device Catalogue Number82406
Device Lot Number2103303330
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/08/2021
Initial Date FDA Received06/08/2021
Supplement Dates Manufacturer Received07/19/2021
Supplement Dates FDA Received08/11/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-