Catalog Number 261221
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem Brain Injury (2219)
Event Date 06/01/2021
Event Type  Injury  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A physician reported a perforator was used on a (b)(6) year-old female patient for a rupture aneurism clipping.The perforator's stopper did not work with 2 burr holes and disengagement failure occurred causing dural damage, minor brain contusion and brain parenchymal surface damage.Treatment with hemostasis was performed.The procedure was completed with a replacement product and no surgical delay was reported.It is unknown the manufacturer for the drill used with the perforator.It is also unknown if the drill was electronic or pneumatic, if the perforator click in place with the drill and if the recommended spring tests were performed between each burr hole.
 
Event Description
N/a.
 
Manufacturer Narrative
Udi: (b)(4).The perforator was not returned for evaluation (as per customer, product not available) and no lot number information has been provided; therefore, an evaluation of the device could not be performed, and manufacturing records could not be reviewed.The cause(s) of the difficulty reported by the customer could not be determined.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOS PERFORATOR
Type of Device
DISPOSABLE PERFORATORS
Manufacturer (Section D)
RAYNHAM
325 paramount drive
325 paramount drive
raynham MA
MDR Report Key11967454
MDR Text Key255153217
Report Number3014334038-2021-00116
Device Sequence Number1
Product Code HBF
Combination Product (y/n)N
PMA/PMN Number
K791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 07/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/09/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number261221
Was Device Available for Evaluation? No
Date Manufacturer Received06/18/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age66 YR
-
-