Related manufacturer reference number: 2017865-2021-21228.It was reported the patient presented remotely.Upon review of the transmission, it was observed the right ventricular lead had low defibrillation impedance.The system was tested in clinic to see normal values, but programming changes were completed to address the impedance value.A year later, the patient presented in clinic having received shock therapy that failed to treat the arrhythmia.The arrhythmia self terminated moments later.An alert was observed for possible lead damage.No further intervention had been completed.The patient was stable.
|