Model Number 0279401200 |
Device Problems
Continuous Firing (1123); Activation, Positioning or Separation Problem (2906)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/04/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Additional information will be provided once the investigation has been completed.
|
|
Event Description
|
It was reported that the device was continuously activating.
|
|
Manufacturer Narrative
|
The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.Alleged failure: probe activated without pushing the button and would not shut off.The probable root cause/s could be an internal leak due to a broken/damaged bond of the polyimide and suction tubing causing fluid ingress to the pcb which can lead to a short circuit.
|
|
Event Description
|
It was reported that the device was continuously activating.
|
|