• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. FLEXIBLE SHAFT W/CIR CONNECTOR; PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. FLEXIBLE SHAFT W/CIR CONNECTOR; PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED Back to Search Results
Model Number 71118200
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/27/2021
Event Type  Injury  
Event Description
It was reported that, during femoral canal reaming of intertan nailing, a flex shaft w/cir connector broke.Surgery was resumed, without any significant delay, with an available smith and nephew back up device; therefore, patient was not harmed.
 
Manufacturer Narrative
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The part of the device that would connect with the corresponding device has broken off, rendering the device inoperative.The device shows signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIBLE SHAFT W/CIR CONNECTOR
Type of Device
PROSTHESIS, SHOULDER, HEMI-, HUMERAL, METALLIC UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11988570
MDR Text Key255872269
Report Number1020279-2021-05129
Device Sequence Number1
Product Code HSD
UDI-Device Identifier03596010436528
UDI-Public03596010436528
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 07/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71118200
Device Catalogue Number71118200
Device Lot Number16JM02366
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/29/2021
Initial Date Manufacturer Received 05/27/2021
Initial Date FDA Received06/12/2021
Supplement Dates Manufacturer Received07/08/2021
Supplement Dates FDA Received07/09/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-