Model Number 6300
Device Problem Improper Flow or Infusion (2954)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Event Description
Information was received indicating that a smiths medical cadd legacy pca pump failed volume accuracy test.There was no reported adverse event.
 
Manufacturer Narrative
Device evaluation: the device was returned for investigation.There was no evidence of the reported problem in the event log.Three separate delivery accuracy tests were performed, and the pump was found to be under delivering to the manufacturing specifications.The root cause of the reported problem cannot be established.An expulsor assembly will be replaced.A dhr review was performed subsequent to the manufacturing of the device and prior to its release.No problems or issues were identified during this dhr review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD LEGACY PCA PUMP
Type of Device
PUMP, INFUSION, PCA
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key11992397
MDR Text Key255973871
Report Number3012307300-2021-05978
Device Sequence Number1
Product Code MEA
UDI-Device Identifier10610586019449
UDI-Public10610586019449
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K982839
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/14/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number6300
Device Catalogue Number21-6300-51
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/28/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received08/06/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/19/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-