ABBOTT VASCULAR G4 STEERABLE GUIDING CATHETER; CATHETER, STEERABLE Back to Search Results
Catalog Number SGC0705
Device Problems Failure to Advance (2524); Deformation Due to Compressive Stress (2889)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/25/2021
Event Type  malfunction  
Manufacturer Narrative
The device will not be returned for evaluation, the device was reportedly discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.The additional device referenced in b5 is being filed under a separate medwatch report number.
 
Event Description
This is filed to report soft tip damage.It was reported that during advancement into the anatomy through the leg, both steerable guide catheters (sgc) could not advance due to the anatomy and the tips became damaged and wrinkled.The anatomy was very tortuous.The procedure was aborted.There were no adverse patient effects and there was no reported clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot-specific quality issue.The investigation determined the reported failure to advance appears to be related to patient morphology/pathology.The damaged and wrinkled soft tip was a cascading effect of the failure to advance; therefore, attributed to procedural condition.There is no indication of a product issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G4 STEERABLE GUIDING CATHETER
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key11995254
MDR Text Key256216339
Report Number2024168-2021-04951
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
PMA/PMN Number
K190167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/14/2022
Device Catalogue NumberSGC0705
Device Lot Number10212U111
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/25/2021
Initial Date FDA Received06/14/2021
Supplement Dates Manufacturer Received07/14/2021
Supplement Dates FDA Received08/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age74 YR
Patient Weight75
-
-