Model Number SGC0701 |
Device Problems
Break (1069); Failure to Advance (2524); Deformation Due to Compressive Stress (2889); Unintended Movement (3026)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/25/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device will not be returned for evaluation, the device was reportedly discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
|
Event Description
|
This is filed to report the suspected cable break.It was reported that this was a mitraclip procedure to treat degenerative mitral regurgitation (mr) with an mr grade of 4.The anatomy was tortuous.The steerable guiding catheter (sgc) failed to cross the septum, it would not pass the iliac due to resistance with the anatomy, and the shaft kinked.The sgc no longer steered when applying minus on the +/- knob.A cable break was suspected.The sgc was removed without issue.A new sgc was used.Two clips were implanted, reducing mr to 1.There were no adverse patient effects and no clinically significant delay during the procedure.No additional information was provided.
|
|
Manufacturer Narrative
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other complaint reported from this lot.All available information was investigated, and a conclusive cause for the reported failure to advance could not be determined in this complaint.The reported deformation due to compressive stress, unintended movement and break were cascading events that resulted from the troubleshooting steps/procedural circumstances.There is no indication of product issue with respect to manufacture, design or labeling.
|
|
Search Alerts/Recalls
|
|
|