• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) UNIFY ASSURA CRT-D RF HV; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) UNIFY ASSURA CRT-D RF HV; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3357-40C
Device Problem Delayed Charge Time (2586)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/27/2021
Event Type  Injury  
Event Description
It was reported that the presented remotely via merlin.Net.Review of the transmission revealed that the implantable cardioverter defibrillator exhibited slow charging.The device was explanted and replaced.There were no patient consequences.
 
Manufacturer Narrative
The reported field event of extended charge time was confirmed in the lab.The device was above the elective replacement indicator (eri) when received.The root cause of the capacitor charge time out could not be conclusively determined.Telemetry, pacing, sensing, impedance, hv output, hv shock, and patient notifier were tested on the bench.No anomalies were detected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY ASSURA CRT-D RF HV
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key12005108
MDR Text Key256451694
Report Number2017865-2021-22199
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734508186
UDI-Public05414734508186
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/15/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2020
Device Model NumberCD3357-40C
Device Catalogue NumberCD3357-40C
Device Lot NumberA000067778
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/14/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/16/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/17/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age68 YR
Patient SexMale
-
-