ETHICON ENDO-SURGERY, LLC. LIGAMAX-5MM ENDO CLIP APPLIER; CLIP, IMPLANTABLE Back to Search Results
Model Number EL5ML
Device Problems Insufficient Information (3190); Firing Problem (4011)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch #: u93m5r.Date of event is 2021.Event day and event month were not reported.(b)(4).Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.A tyvek was returned along with the device.Upon visual analysis of the returned sample, it was determined that the el5ml device was received with one jaw disengaged from the cam, and the cam was noted bent from the right side.This condition would not allow the jaws to collapse in order to form the clips.No functional test was performed due the condition of the device.In order to evaluate the condition of the internal components of the device, it was disassembled.Upon disassembly, the advancer was noted bent and six clips were found inside clip track.The event reported was confirmed and it is related to improper use of the device.Please note that prior to loading a clip in the jaws, ensure that the demarcation between the jaws and the device shaft is past the end of the trocar cannula.Additionally, excessively applying a side load to the jaws, causing them to partially collapse, could result in a clip malformation.The device jaws should be fully open and parallel upon initiating the firing of the device.Possible causes for the condition of the jaw disengagement from the cam may be due to inadvertent force, twisting, or pressure being placed on the device jaws, using the jaws of the device as a dissector/retractor, or damage to the jaws while entering the trocar.Please reference the instructions for use for more information.As part of our quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.A manufacturing record evaluation was performed for the finished device lot/batch number, and no non-conformances were identified.
 
Event Description
It was reported that during an unknown procedure, the el5ml was hard to fire and clips deflected.It is unknown how the procedure was completed.There was no patient consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGAMAX-5MM ENDO CLIP APPLIER
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo
*  
Manufacturer Contact
kara ditty-bovard
475 calle c
guaynabo 
*  
6107428552
MDR Report Key12005243
MDR Text Key256526432
Report Number3005075853-2021-03342
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036001843
UDI-Public10705036001843
Combination Product (y/n)N
PMA/PMN Number
K050344
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 05/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEL5ML
Device Catalogue NumberEL5ML
Device Lot NumberU93M5R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/20/2021
Initial Date Manufacturer Received 05/20/2021
Initial Date FDA Received06/15/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/11/2020
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-