Model Number 777000
Device Problems Output below Specifications (3004); Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/25/2021
Event Type  malfunction  
Manufacturer Narrative
The returned device was evaluated at olympus australia site.Evaluation found the reported issue was able to be duplicated.The device upon inspection found old version pb board causing output issue.The pb board was replaced and once completed, the device was tested and passed all required testing and specifications.Based on evaluation findings the reported failure was due to old version pb board.The pb board was attributed to component failure.Investigation is ongoing.This report will be supplemented accordingly following investigations.
 
Event Description
It was reported that during device maintenance low pbi output power was observed.There was no patient involvement, no user injury reported due to the event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.All records indicated that the product was manufactured according to all applicable procedures and met final product release criteria.No abnormalities were found.Olympus will continue to monitor complaints for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G400 GENERATOR, GYRUS
Type of Device
G400 GENERATOR, GYRUS
Manufacturer (Section D)
GYRUS ACMI, INC
136 turnpike road
southborough MA 01772
MDR Report Key12005317
MDR Text Key257393122
Report Number3003790304-2021-00105
Device Sequence Number1
Product Code GEI
UDI-Device Identifier00821925009004
UDI-Public00821925009004
Combination Product (y/n)N
PMA/PMN Number
K050550
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 07/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number777000
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/08/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-