Model Number 777000 |
Device Problems
Output below Specifications (3004); Output Problem (3005)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/25/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The returned device was evaluated at olympus australia site.Evaluation found the reported issue was able to be duplicated.The device upon inspection found old version pb board causing output issue.The pb board was replaced and once completed, the device was tested and passed all required testing and specifications.Based on evaluation findings the reported failure was due to old version pb board.The pb board was attributed to component failure.Investigation is ongoing.This report will be supplemented accordingly following investigations.
|
|
Event Description
|
It was reported that during device maintenance low pbi output power was observed.There was no patient involvement, no user injury reported due to the event.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.All records indicated that the product was manufactured according to all applicable procedures and met final product release criteria.No abnormalities were found.Olympus will continue to monitor complaints for this device.
|
|