Model Number 744000
Device Problem No Device Output (1435)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The evaluation of the device found no output power malfunction at the rv1 due to faulty circuit boards.The generator was repaired to asset specifications and returned to asset stock.The asset was last serviced on (b)(6) 2021; no problems were found, inspection only.The investigation is ongoing; therefore, the root cause of the reported malfunction cannot be determined at this time.However, if additional information becomes available, this report will be supplemented accordingly.
 
Event Description
The plasma kinetic superpulse generator was returned to the service center for an asset inspection.During testing, the generator was found to have a no output power malfunction due to faulty circuit boards.There was no reported allegation of a malfunction associated with the device and there was no patient involvement reported.
 
Manufacturer Narrative
This supplemental report was submitted to provide additional information from the legal manufacturer.The legal manufacturer performed the device history records for this device and all records indicated that the product was manufactured according to all applicable procedures and met final product release criteria.No abnormalities were found.The investigation was completed by the legal manufacturer and determined that there is no manufacturing, material or processing related cause for this failure mode.Based on the physical evaluation, the cause of the reported event was attributed to a faulty boards (psu & pkrf) as the boards were replaced and the unit passed all functional tests afterwards.The legal manufacturer will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GYRUS, PK-SP GENERATOR
Type of Device
PK-SP GENERATOR
Manufacturer (Section D)
GYRUS ACMI, INC
136 turnpike road
southborough MA 01772
MDR Report Key12006122
MDR Text Key256556146
Report Number3003790304-2021-00106
Device Sequence Number1
Product Code GEI
UDI-Device Identifier00821925009271
UDI-Public00821925009271
Combination Product (y/n)N
PMA/PMN Number
K031085
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 06/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number744000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/11/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/20/2021
Initial Date FDA Received06/15/2021
Supplement Dates Manufacturer Received06/25/2021
Supplement Dates FDA Received06/28/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-