This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, d4, d9, g3, g6, h1, h2, h3, h4, h6, h10 one g7 10 deg e1 liner 36mm f item# 010000897 lot# 6105189 was returned and evaluated.Upon visual inspection there were two locations where the locking feature had been damaged.There is also an indentation to the outside radius.Medical records were not provided.A review of the device history records identified no deviations or anomalies during manufacturing.A review of complaint history identified additional similar complaints for the reported item and no additional complaints for the reported part and lot combination.Complaints are monitored through monthly complaint review (reference sop040001 and wi040015) in order to identify potential adverse trends.A definitive root cause cannot be determined.A summary of the investigation has been sent to the complainant.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|