ZIMMER BIOMET, INC. G7 10 DEG E1 LINER 36MM F; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Difficult to Insert (1316)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/27/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Part: 010000664, lot: 6405754.Report source: (b)(6).
 
Event Description
It was reported that the g7 poly liner would not mate with the g7 pps acetabular cup during the surgery.The surgery was completed with another liner.No further information is available at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, d4, d9, g3, g6, h1, h2, h3, h4, h6, h10 one g7 10 deg e1 liner 36mm f item# 010000897 lot# 6105189 was returned and evaluated.Upon visual inspection there were two locations where the locking feature had been damaged.There is also an indentation to the outside radius.Medical records were not provided.A review of the device history records identified no deviations or anomalies during manufacturing.A review of complaint history identified additional similar complaints for the reported item and no additional complaints for the reported part and lot combination.Complaints are monitored through monthly complaint review (reference sop040001 and wi040015) in order to identify potential adverse trends.A definitive root cause cannot be determined.A summary of the investigation has been sent to the complainant.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 10 DEG E1 LINER 36MM F
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key12008027
MDR Text Key256458535
Report Number0001825034-2021-01806
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
PMA/PMN Number
K121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/02/2022
Device Model NumberN/A
Device Catalogue Number010000897
Device Lot Number6105189
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/01/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/28/2021
Initial Date FDA Received06/16/2021
Supplement Dates Manufacturer Received09/29/2021
Supplement Dates FDA Received10/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-