The following was reported to gore: on (b)(6) 2020, the patient underwent a norwood procedure utilizing a gore-tex® vascular graft configured for pediatric shunt device.Physician confirmed all shunts are on prophylactic heparin until they¿re able to transition to aspirin and that aspirin responsiveness is checked in everyone prior to stopping heparin.On (b)(6) 2020, the shunt is found to be patent but mildly narrowed at its proximal anastomosis.A heparin drip is started on the patient.On (b)(6) 2020 a cardiac catheter with a stent is placed in the gore-tex® vascular graft configured for pediatric shunt.On (b)(6) 2020 an echo shows the pediatric shunt is patent but there is less flow through the shunt when compared to the previous echo which could indicate proximal or distal narrowing.On (b)(6) 2020, shunt ligation and takedown is performed and a "yellowish fluid" is found around the shunt.Samples from the yellowish fluid produced no growth on cultures.
|