MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
|
Back to Search Results |
|
Model Number 37603 |
Device Problems
Break (1069); High impedance (1291); Migration or Expulsion of Device (1395); Inappropriate/Inadequate Shock/Stimulation (1574)
|
Patient Problems
Muscular Rigidity (1968); Shaking/Tremors (2515); Electric Shock (2554)
|
Event Date 06/16/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a healthcare provider (hcp), via the manufacturer¿s representative (rep), who reported the patient said they kept getting ¿wrong electrical shocks and muscles tightness on their right side throughout the day.There were no recent falls noted.An impedance check was performed and contact 0 showed an open circuit which was the contact being used for therapy.The device was programmed using contact 1 which seemed to help their tremor resolving the issue.Relevant medical history includes parkinson¿s disease.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received from a manufacturer representative (rep) reporting that impedances were rechecked because tremor returned and all 4 contacts were now open.The patient is not receiving therapy at this time.The patient denies any falls or trauma to the device.X-ray showed that the lead is fractured just above the extension connection.It appears that the lead/extension connection area slid down the neck over time, which could be the cause of the lead fracture.The patient will be scheduled for a lead replacement in the near future.
|
|
Manufacturer Narrative
|
Other relevant device(s) are: product id: 3389s-40, lot#: va1nelu, implanted: (b)(6) 2018, product type: lead.Product id: 3708660, serial#: (b)(6), product type: extension.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|