Model Number GZ-130PA |
Device Problem
Communication or Transmission Problem (2896)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/20/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The customer reported that there was momentary interruption in patient monitoring, the transmitter will not power up.The customer will send in the unit for exchange.There was no patient injury reported.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
|
|
Event Description
|
The customer reported that the gz monitor had momentary interruption in patient monitoring.
|
|
Manufacturer Narrative
|
The customer reported that the gz transmitter had a momentary interruption in patient monitoring.The transmitter was then failing to power up.No patient harm was reported.Investigation summary: the device was sent in for evaluation.Startup failure was confirmed during the nihon kohden repair center (nk rc) evaluation.Fluid intrusion was also found in the unit.The cause of startup failure is likely related to fluid intrusion causing damage to the electrical components.Fluid intrusion is likely a result of mishandling of the device by the user.Additional information: b4 date of this report.D9 device available for evaluation? g3 date received by manufacturer.G6 type of report.H2 if follow-up, what type? h3 device evaluated by manufacturer? h6 event problem and evaluation codes.H10 additional manufacturer narrative.
|
|
Event Description
|
The customer reported that the gz transmitter had a momentary interruption in patient monitoring.No patient harm was reported.
|
|
Search Alerts/Recalls
|