• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NIHON KOHDEN CORPORATION GZ-130PA; TRANSMITTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN CORPORATION GZ-130PA; TRANSMITTER Back to Search Results
Model Number GZ-130PA
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/20/2021
Event Type  malfunction  
Manufacturer Narrative
The customer reported that there was momentary interruption in patient monitoring, the transmitter will not power up.The customer will send in the unit for exchange.There was no patient injury reported.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
 
Event Description
The customer reported that the gz monitor had momentary interruption in patient monitoring.
 
Manufacturer Narrative
The customer reported that the gz transmitter had a momentary interruption in patient monitoring.The transmitter was then failing to power up.No patient harm was reported.Investigation summary: the device was sent in for evaluation.Startup failure was confirmed during the nihon kohden repair center (nk rc) evaluation.Fluid intrusion was also found in the unit.The cause of startup failure is likely related to fluid intrusion causing damage to the electrical components.Fluid intrusion is likely a result of mishandling of the device by the user.Additional information: b4 date of this report.D9 device available for evaluation? g3 date received by manufacturer.G6 type of report.H2 if follow-up, what type? h3 device evaluated by manufacturer? h6 event problem and evaluation codes.H10 additional manufacturer narrative.
 
Event Description
The customer reported that the gz transmitter had a momentary interruption in patient monitoring.No patient harm was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GZ-130PA
Type of Device
TRANSMITTER
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
attn: shama mooman
1-31-4 nishiochia
shinjuku-ku, tokyo 116-8 560
JA  116-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
attn: shama mooman
1-1 tajino
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
seibu bldg 2, 4th floor 1-11-2
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key12021438
MDR Text Key280788623
Report Number8030229-2021-00352
Device Sequence Number1
Product Code DRT
UDI-Device Identifier04931921117415
UDI-Public04931921117415
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153707
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 09/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGZ-130PA
Device Catalogue NumberGZ-130PA
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/20/2021
Initial Date FDA Received06/17/2021
Supplement Dates Manufacturer Received08/23/2023
Supplement Dates FDA Received09/06/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/11/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
CNS; CNS
-
-