MAUDE Adverse Event Report: VENUSA DE MEXICO S.A. DE C.V. TACTICATH QUARTZ CONTACT FORCE ABLATION CATHETER, 75MM; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VENUSA DE MEXICO S.A. DE C.V. TACTICATH QUARTZ CONTACT FORCE ABLATION CATHETER, 75MM; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number PN-004 075
Device Problems Fluid/Blood Leak (1250); Optical Problem (3001)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/08/2021
Event Type  malfunction  
Manufacturer Narrative
One 75 mm tacticath quartz contact force ablation catheter was received for evaluation.The device did not meet specifications during a shaft leak test and an irrigation leak test.Further investigation revealed the shaft material had been fractured at the distal edge of electrode ring 4, consistent with the failed shaft and irrigation leak tests, fluid ingress proximal to electrode ring 4, optical signal interruption in fibers 1-3, and a loss of force data during the procedure.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the damaged shaft material, shaft leak, and fractured and exposed conductor wire was unable to be determined and remains unknown.
 
Event Description
This report is to advise of an event observed during analysis confirming a shaft leak.
 
Manufacturer Narrative
One 75 mm tacticath quartz contact force ablation catheter was received for evaluation.Optical fibers 1-3 no longer met specifications for optical signal properties and no contact force was displayed when connected to the tactisys quartz unit.Further investigation revealed that optical fibers 1-3 were under-polished, resulting in the obstruction of the optical fiber core, consistent with the observed signal loss and with the reported event.In addition, an adhesive-like substance was confirmed in the optical fiber cavity of fibers 1-3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TACTICATH QUARTZ CONTACT FORCE ABLATION CATHETER, 75MM
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
VENUSA DE MEXICO S.A. DE C.V.
calle hertz 1525-6
parque industrial j. bermudez
ciudad juarez, chihuahua 32470
MX  32470
MDR Report Key12031458
MDR Text Key257132453
Report Number9680001-2021-00040
Device Sequence Number1
Product Code OAE
UDI-Device Identifier07640157990040
UDI-Public07640157990040
Combination Product (y/n)N
PMA/PMN Number
P130026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/21/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/25/2022
Device Model NumberPN-004 075
Device Catalogue NumberPN-004075
Device Lot Number7300931
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/04/2021
Was the Report Sent to FDA? No
Date Manufacturer Received09/10/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-