VENUSA DE MEXICO S.A. DE C.V. TACTICATH QUARTZ CONTACT FORCE ABLATION CATHETER, 75MM; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
|
Back to Search Results |
|
Model Number PN-004 075 |
Device Problems
Fluid/Blood Leak (1250); Optical Problem (3001)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/08/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
One 75 mm tacticath quartz contact force ablation catheter was received for evaluation.The device did not meet specifications during a shaft leak test and an irrigation leak test.Further investigation revealed the shaft material had been fractured at the distal edge of electrode ring 4, consistent with the failed shaft and irrigation leak tests, fluid ingress proximal to electrode ring 4, optical signal interruption in fibers 1-3, and a loss of force data during the procedure.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the damaged shaft material, shaft leak, and fractured and exposed conductor wire was unable to be determined and remains unknown.
|
|
Event Description
|
This report is to advise of an event observed during analysis confirming a shaft leak.
|
|
Manufacturer Narrative
|
One 75 mm tacticath quartz contact force ablation catheter was received for evaluation.Optical fibers 1-3 no longer met specifications for optical signal properties and no contact force was displayed when connected to the tactisys quartz unit.Further investigation revealed that optical fibers 1-3 were under-polished, resulting in the obstruction of the optical fiber core, consistent with the observed signal loss and with the reported event.In addition, an adhesive-like substance was confirmed in the optical fiber cavity of fibers 1-3.
|
|
Search Alerts/Recalls
|
|
|