Catalog Number 261221 |
Device Problem
Material Separation (1562)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/01/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Attempts are being made to obtain additional information.Upon completion of the investigation, a follow-up report will be submitted.
|
|
Event Description
|
A facility reported that during use, the codman disposable perforator (261221) came apart in the patient's head.No patient injury or surgical delay was reported.
|
|
Manufacturer Narrative
|
The perforator was returned for evaluation.Device history record (dhr) - there is no indication that the production process may have contributed to this complaint.All test results passed procedural specifications.Failure analysis - the perforator unit was inspected using the unaided eye.Unit was lightly soiled and was received disassembled.Ifu testing procedure was performed with no observed anomalies after being reassembled/resleeved.Functional testing was performed using the same protocol it underwent at finished goods testing prior to release.The unit was found to perform as intended and fulfilled the acceptance criteria.In the failure analysis that was performed, the returned unit was found to work as intended, and met all acceptance criteria.The complaint could not be verified through failure analysis.The root cause is undetermined and was unable to be confirmed in the complaint evaluation.Product was received for analysis and the investigation could not confirm the complaint.
|
|
Event Description
|
N/a.
|
|