During an in clinic follow-up, excessive bluetooth use was detected on the device.Interrogation was attempted but the device was unable to communicate via bluetooth telemetry.However, the device was able to be interrogated using inductive telemetry.The patient was in stable condition.
|
The reported event of unable to communicate via bluetooth telemetry and lockout were confirmed.The logs were reviewed and determined there was a memory corruption which resulted in connection errors.The connection errors led to a cyber lockout, however, this does not affect device therapy functions.
|