• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT NEUTRINO NXT DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT NEUTRINO NXT DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CDDRA600Q
Device Problems Failure to Interrogate (1332); Wireless Communication Problem (3283)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/07/2021
Event Type  malfunction  
Event Description
During an in clinic follow-up, excessive bluetooth use was detected on the device.Interrogation was attempted but the device was unable to communicate via bluetooth telemetry.However, the device was able to be interrogated using inductive telemetry.The patient was in stable condition.
 
Manufacturer Narrative
The reported event of unable to communicate via bluetooth telemetry and lockout were confirmed.The logs were reviewed and determined there was a memory corruption which resulted in connection errors.The connection errors led to a cyber lockout, however, this does not affect device therapy functions.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUTRINO NXT DR
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
MDR Report Key12046035
MDR Text Key257734736
Report Number2017865-2021-23804
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2022
Device Model NumberCDDRA600Q
Device Lot NumberS000079138
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/07/2021
Initial Date FDA Received06/22/2021
Supplement Dates Manufacturer Received09/10/2021
Supplement Dates FDA Received09/22/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-