• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEOTRACT, INC. NEOTRACT UROLIFT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEOTRACT, INC. NEOTRACT UROLIFT SYSTEM Back to Search Results
Model Number UROLIFT SYSTEM
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Failure of Implant (1924)
Event Date 05/25/2021
Event Type  Injury  
Event Description
On (b)(6) 2021, neotract became aware of a patient whose inflatable penile prosthesis (ipp) would not inflate following a prostatic urethral lift (pul) procedure.On (b)(6) 2021, the physician removed and replaced the ipp and noted that the urolift implant suture was deployed inside the ipp reservoir.The patient was reported to be doing well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEOTRACT UROLIFT SYSTEM
Type of Device
UROLIFT SYSTEM
Manufacturer (Section D)
NEOTRACT, INC.
4155 hopyard road
pleasanton CA 94588
Manufacturer (Section G)
NEOTRACT, INC.
4155 hopyard road
pleasanton CA 94588
Manufacturer Contact
brian gall
4155 hopyard road
pleasanton, CA 94588
9253296547
MDR Report Key12046649
MDR Text Key260757026
Report Number3015181082-2021-00017
Device Sequence Number1
Product Code PEW
UDI-Device Identifier00814932020001
UDI-Public00814932020001
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K193269
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 06/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/22/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberUROLIFT SYSTEM
Device Catalogue NumberAN00155
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/25/2021
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-