SMITH & NEPHEW, INC. UNKN JOURNEY BCS / JOURNEY II BCS KNEE FEM COMP; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number UNKN01200303
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Unspecified Infection (1930); Joint Laxity (4526); Implant Pain (4561)
Event Date 06/01/2021
Event Type  Injury  
Event Description
It was reported that, after knee replacement surgery with a journey ii bcs system had been performed on (b)(6) 2018, the patient experienced pain, loosening and an infection.This adverse event was treated by a revision surgery on (b)(6) 2021.All four components, insert, femur, tibia, and patella, were explanted.Current health status of patient is unknown.
 
Manufacturer Narrative
H3, h6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, without the requested clinical information, the implantation/revision reports, lab findings, radiographs or the device, the clinical root cause of the reported pain, loosening and an infection cannot be determined.Although it was reported this adverse with was treated with a revision with the removal of the insert, femur, tibia, and patella.The impact to the patient beyond that which has already been reported cannot be determined.Should any additional medical information be provided, this complaint will be re-assessed.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to contamination, post-operative healing issue, abnormal motion over time, bone degeneration, fit/sizing, lack of ingrowth, lifetime of device, joint tightness, material in use, patient reaction and/or traumatic injury.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Additional information: d3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKN JOURNEY BCS / JOURNEY II BCS KNEE FEM COMP
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12059860
MDR Text Key258119877
Report Number1020279-2021-05429
Device Sequence Number1
Product Code HRY
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 07/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKN01200303
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/11/2021
Initial Date FDA Received06/24/2021
Supplement Dates Manufacturer Received07/21/2021
Supplement Dates FDA Received07/22/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age60 YR
Patient Weight75
-
-