• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION NS FX25REC W/RES; BLOOD GAS OXYGENATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION NS FX25REC W/RES; BLOOD GAS OXYGENATOR Back to Search Results
Model Number 3ZZ*FX25RECA
Device Problem Coagulation in Device or Device Ingredient (1096)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/07/2021
Event Type  malfunction  
Manufacturer Narrative
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.
 
Event Description
The user facility reported to terumo cardiovascular that after cardiopulmonary bypass, the oxygenator clotted.No patient involvement.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on june 25, 2021.Upon further investigation of the reported event, the following information is new and/or changed: d9 (device availability - added date returned to manufacturer) g3 (date received by manufacturer) g6 (indication that this is a follow-up report) h2 (follow-up due to additional information) h3 (device evaluation anticipated by manufacturer - a second follow-up will be submitted upon completion of the investigation and/or submission of new information, thus tcvs references conclusions code 11.) all available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations.Type of investigation: #1:10 - testing of actual/suspected device.Type of investigation #2: 11 - testing of device from same lot/batch retained by manufacturer.Type of investigation #3: 3331 - analysis of production records.Investigation findings: 213 - no device problem found.Investigation conclusions: 67 - no problem detected.The sample was visually inspected and did not find any anomaly including a breakage.It was installed into a circuit consisting of tubes, and the pressure drop when the circulating bovine blood (hct: 35%, temp 37 celsius) was measured.It was confirmed to meet the factory's specifications and no obstruction was found.The water was flushed through the blood channel.No clot was observed.Review of the manufacturing record and the incoming inspection record of the actual sample confirmed that there was no anomaly in them.A search of the complaint file found no other similar report with the involved product code/lot# combination.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NS FX25REC W/RES
Type of Device
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
MDR Report Key12070542
MDR Text Key260451384
Report Number1124841-2021-00150
Device Sequence Number1
Product Code DTZ
Combination Product (y/n)N
PMA/PMN Number
K151791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 08/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2023
Device Model Number3ZZ*FX25RECA
Device Catalogue NumberN/A
Device Lot NumberYN16
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/09/2021
Initial Date FDA Received06/25/2021
Supplement Dates Manufacturer Received06/28/2021
08/12/2021
Supplement Dates FDA Received07/16/2021
08/21/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-