Pump s/n: (b)(4) was received and could not be tested to confirm the reported issue due to physical damage.During visual evaluation it was noted that the gearbox assembly was damaged, the rotor assembly was stuck and the back case was cracked.The damaged parts were replaced and the pump was tested and passed all functional tests.The service history record was reviewed.This device was manufactured in 2013 and this is the first time this device has been returned for service.The damaged components were determined to be caused by customer mishandling of the device.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.
|