• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN KANGAROO JOEYPUMP W/POLE CLMP; PUMP, INFUSION, ENTERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN KANGAROO JOEYPUMP W/POLE CLMP; PUMP, INFUSION, ENTERAL Back to Search Results
Model Number 383400
Device Problem Excess Flow or Over-Infusion (1311)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The returned pump s/n: (b)(4) was evaluated.The pump was received in a used condition.The pump was functionally tested and the customer reported issue was not confirmed.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.
 
Event Description
The customer reported that the device overfeeds.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO JOEYPUMP W/POLE CLMP
Type of Device
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
COVIDIEN
15 hampshire street
mansfield MA 02048
Manufacturer (Section G)
COVIDIEN
15 hampshire street
mansfield MA 02048
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key12072924
MDR Text Key258968414
Report Number1282497-2021-10448
Device Sequence Number1
Product Code LZH
UDI-Device Identifier10884521006645
UDI-Public10884521006645
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 06/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/26/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number383400
Device Catalogue Number383400
Was Device Available for Evaluation? Yes
Date Manufacturer Received01/10/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Type of Device Usage N
Patient Sequence Number1
-
-