• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WARSAW ORTHOPEDICS VERTEX® RECONSTRUCTION SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WARSAW ORTHOPEDICS VERTEX® RECONSTRUCTION SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL Back to Search Results
Model Number G7752536
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/03/2021
Event Type  malfunction  
Manufacturer Narrative
This part is not approved for use in the united states; however, a like device catalog # 7752535, 510k #k082728, udi#: (b)(4) was cleared in the united states.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from the healthcare professional (hcp) via the manufacturer representative regarding an event which occurred during an c4-th2 posterior fusion procedure in a patient diagnosed with c6 dislocation fracture.It was reported that after placing crosslink on c6 and temporarily tightening it the set screw of the rod crosslink stripped during the final tightening, so it was replaced with a new crosslink (same size).There was a delay of less than 60 mins in the overall procedure.No patient injury / complication was reported.
 
Manufacturer Narrative
H3: part no.: g7752536 lot# 0807676w visual and optical examination revealed the female hex edges have been rounded/deformed.There were no signs of thread damage on the set screw.The threads appear to be used but no signs of cross threading.The damage to the hex appears to be from torsional overload.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERTEX® RECONSTRUCTION SYSTEM
Type of Device
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
Manufacturer (Section G)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key12074182
MDR Text Key265805966
Report Number1030489-2021-00850
Device Sequence Number1
Product Code KWP
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
SEE H10
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG7752536
Device Catalogue NumberG7752536
Device Lot Number0807676W
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/17/2021
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/03/2021
Initial Date FDA Received06/28/2021
Supplement Dates Manufacturer Received11/23/2021
Supplement Dates FDA Received12/15/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/20/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-