• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE GMBH ARTIS ZEE FLOOR; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE GMBH ARTIS ZEE FLOOR; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number 10094135
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/21/2021
Event Type  malfunction  
Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Event Description
It was reported to siemens that a malfunction occurred while operating the artis zee floor system.During an interventional procedure, the user reported that the collimator was not aligned.The procedure was continued and completed on an alternate system.We are unaware of any impact to the state of health of the patient involved.Siemens has requested additional information in order to conduct an investigation of the reported event.
 
Manufacturer Narrative
Siemens has completed an investigation of the reported event.The root cause was determined to be a hardware error.The investigation could not provide a clear result although it was determined that there were problems with the collimator on site.Conceivable here is a temporary contact problem that was resolved by the service activities.The collimator was replaced, and no error could be detected when the returned part was examined.The error described did not lead to a complete loss of x-ray radiation, but only to the fact that the system generated an error message "collimator not aligned" and the flat panel detector could not be rotated or aligned by motor.The collimator was replaced and the system then functioned again as specified.The spare parts consumption was checked and a possible accumulation of faults or even a possible general fault that would require corrective measures of the installed base could not be determined by the investigation.After detailed investigation, the incident is not classified as a reportable event as neither serious injury, death nor an unexpected, prolonged hospitalization of the patient or any other person occurred or could be expected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTIS ZEE FLOOR
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
SIEMENS HEALTHCARE GMBH
siemensstr. 1
forchheim, germany 91301
GM  91301
MDR Report Key12074845
MDR Text Key258654289
Report Number3004977335-2021-85939
Device Sequence Number1
Product Code OWB
UDI-Device Identifier04056869010045
UDI-Public04056869010045
Combination Product (y/n)N
PMA/PMN Number
K181407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Repair
Type of Report Initial,Followup
Report Date 06/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10094135
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 06/21/2021
Initial Date FDA Received06/28/2021
Supplement Dates Manufacturer Received08/27/2021
Supplement Dates FDA Received08/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-