This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.The following sections were updated/corrected: b4, b5, d4, g3, g6, h2, h3, h4, h6, h8, h10.Review of the most recent repair record determined the unit was out of calibration, the control bar had thin ends, and a screw was broken into the machined head.The shaft bearings, sleeve bearings, control bar, machined head and screws were replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.Device is used for treatment.A definitive root cause cannot be determined.The event is confirmed.
|