Model Number 8888145001 |
Device Problem
Obstruction of Flow (2423)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/10/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during the insertion procedure, the product was defective blocked in one lumen.There catheter was not repaired and 2% chlorohexidine was used as the cleaning agent on the device.Tego was utilized.There was no leak and there was no luer adapter issue.An unknown solution was used on the insertion site prior to product placement.There was nothing unusual observed on the device prior to use and flushing was not done.There were no other products being utilized with the device and there were no other defects/damages found on the product.There was no blood loss.A new product was used to resolve the issue and the procedure was completed.There was no patient injury.
|
|
Manufacturer Narrative
|
Additional information: b5, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the device was occluded.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during the insertion procedure, the product was defective (blocked in one lumen).The catheter was not repaired and 2% chlorohexidine was used as the cleaning agent on the device.Tego was utilized.There was no leak and there was no luer adapter issue.An unknown solution was used on the insertion site prior to product placement.There was nothing unusual observed on the device prior to use and flushing was not done.There were no other products being utilized with the device and there were no other defects/damages found on the product.The customer did not try to reverse the lines.There was no blood loss.The product was replaced immediately with a new one from the hospital's stock to resolve the issue and the procedure was completed.There was no patient injury.
|
|
Event Description
|
According to the reporter, during the insertion procedure, the product was defective (blocked in one lumen).There catheter was not repaired and 2% chlorohexidine was used as the cleaning agent on the device.Tego was utilized.There was no leak and there was no luer adapter issue.An unknown solution was used on the insertion site prior to product placement.There was nothing unusual observed on the device prior to use and flushing was not done.There were no other products being utilized with the device and there were no other defects/damages found on the product.The customer did not try to reverse the lines.There was no blood loss.The product was replaced immediately with a new one from the hospital's stock to resolve the issue and the procedure was completed.There was no patient injury.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|