Model Number PM3242 |
Device Problems
Failure to Disconnect (2541); Contamination of Device Ingredient or Reagent (2901)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/24/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
Related manufacturer reference number: 2017865-2021-24271.Related manufacturer reference number: 2017865-2021-24272.It was reported that the patient presented in clinic for a generator change procedure.During the procedure, the right atrial (ra) and right ventricular (rv) leads could not be removed due to tissue in the header of the pacemaker.Additionally, the ra and rv leads insulation was damaged during the procedure.The pacemaker was removed and replaced.Both leads were capped and the rv lead was replaced on (b)(6) 2021.The patient had no adverse consequences.
|
|
Manufacturer Narrative
|
H6 should include 2541 - failure to disconnect.
|
|
Manufacturer Narrative
|
The reported inability to remove the leads from the header was not confirmed in the laboratory.The device was found to have the original sbp (scalable brady platform) header design which is not compatible with lead connectors.The root cause is associated with the is-1 bore in sbp models having dimensions that are smaller in certain areas.The cause of the inability to remove leads was the header design.
|
|
Search Alerts/Recalls
|