It was reported that the ipg would not connect to the clinician programmer post implant.Service app mode, a safety mode enabled by the infinity ipg, was enabled.The clinician programmer displayed "cannot connect to generator.All appropriate trouble shooting steps were performed.Ipg was unable to be recovered and was deemed to be inoperable.The patient was thereby taken again to the operating room and the ipg was replaced with a new one.This addressed the issue.
|
The device is included in the neuromodulation implantable pulse generator (ipg) inoperable when exposed to monopolar electrosurgery advisory notice issued by abbott on 02 june 2017.The reported event for no ipg communication was confirmed.It was determined the device was running the service application software.This condition is consistent with electrocautery use during surgery.Per event details, the ipg no longer communicated following the ipg implant procedure.There is guidance in the clinicians manual regarding proper handling of the device when using electrocautery.As a result of this finding, actions have been taken to prevent reoccurrence.
|