Model Number 170053 |
Device Problems
Detachment of Device or Device Component (2907); Human-Device Interface Problem (2949)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/07/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during laparoscopic roux en y, when closing gastrojejunostomy, the device worked well for the first few suture passes, then the needle got caught up in tissue and popped off into the patient's cavity.Once the surgeon re-attached the needle, it was difficult to pass the needle made it work.The nurse was unable to unload the device.Another handle and reload was used to resolve the issue in order to complete the case.There was no patient injury.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during laparoscopic rouxeny, when closing gastrojejunostomy, the device worked well for the first few suture passes, then the needle got caught up in tissue and popped off from the suture into the patient's cavity.A laparoscopic grasper was used to retrieve the component.Once the surgeon re-attached the needle, it was difficult to pass the needle made it work.The nurse was unable to unload the device.Another handle and reload was used to resolve the issue in order to complete the case.There was no patient injury.
|
|
Search Alerts/Recalls
|
|