• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HOL ML 5MM ENDO APPLIER; APPLIER, SURGICAL, CLIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HOL ML 5MM ENDO APPLIER; APPLIER, SURGICAL, CLIP Back to Search Results
Catalog Number 544965
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/21/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event Description
The pivot pin got detached during a surgery.Therefore, the applier was replaced with a new one.Nothing fell/remained in the patient.The applier was purchased by the hospital in (b)(6) 2020.
 
Manufacturer Narrative
Qn#(b)(6) the dhr for the returned instrument was reviewed and found completely without any irregularities.This instrument was produced at the tecomet , inc.Kenosha wi facility as part of a (b)(4) pc.Lot in february of 2020.Evaluation of the returned instrument shows that the tips are slightly loose and misaligned to each other in the closed position.There was no visible damage found to the jaw pivot pin as stated in this complaint.After the initial evaluation this instrument was dis-assembled in order to evaluate its internal components and it was found that the inner drive rod (n00185) fingers are both damaged where they engage the jaws.It is suspected that the damaged drive rod fingers caused the jaws to become slightly loose and misaligned in the closed position.We are unable to determine what caused the drive rod fingers to become damaged but mishandling of this device at the end user's facility is suspected.All (b)(4) instruments from this lot were 100% visually inspected and function tested prior to shipment to the customer as this is a standardized procedure at this facility for this product line.Due to these findings, no further actions will be taken in response to this complaint and this record will be deemed closed.
 
Event Description
The pivot pin got detached during a surgery.Therefore, the applier was replaced with a new one.Nothing fell/remained in the patient.The applier was purchased by the hospital in october 2020.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOL ML 5MM ENDO APPLIER
Type of Device
APPLIER, SURGICAL, CLIP
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
MDR Report Key12085116
MDR Text Key258950157
Report Number3011137372-2021-00162
Device Sequence Number1
Product Code GDO
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 06/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number544965
Device Lot Number06F1989607
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/09/2021
Initial Date Manufacturer Received 06/21/2021
Initial Date FDA Received06/29/2021
Supplement Dates Manufacturer Received07/26/2021
Supplement Dates FDA Received07/27/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-