Device history record review: the dhr shows no abnormalities related to the reported failure.Mayfield modified skull clamp (a1059) was returned for evaluation.The evaluation of the device could not duplicate the reported complaint.With respect to the returned unit it has passed all specific functional testing requirements, except for the lock having rotational movement when unit is not under pressure, this would not have caused a slippage; when unit is properly positioned and put under pressure unit would not have slipped.General maintenance and cleaning required at this time.The reported complaint is not confirmed form the evaluation.No further investigation required based on the acceptability of risk and no adverse trends identified.This will be monitored and trended going forward.
|
Medwatch form uf/ importer report # (b)(4) was received on 03jun2021 in relation to mayfield modified skull clamp (a1059): a (b)(6) year old female patient underwent a craniotomy procedure on (b)(6) 2021.The neuro resident pinned the patient's head using the mayfield c clamp.The patient was then positioned prone onto the jackson table.The neuro resident was holding the c clamp attached to the patient's head and while flexing the patient's head, there was a clicking sound and the patient's head moved.A laceration was then noted on the patient's left head near one of the mayfield pins.A neuro physician came into the room to assess and did education of pinning to the neuro resident.Patient was positioned back onto cart, and mayfield c clamp was reapplied.The attending physician assessed, shaved, cleaned up, and stapled laceration.Patient was then positioned back onto the jackson table.
|