• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO GASTROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO GASTROSCOPE Back to Search Results
Model Number EG16-K10
Device Problems Failure to Disconnect (2541); Device Damaged by Another Device (2915)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/25/2021
Event Type  malfunction  
Event Description
Pentax medical became aware of an event on 25-feb-2021 that occurred in (b)(6).The reported complaint of "forceps mouth brush clogging", involving pentax medical video gastroscope model eg16-k10, serial number (b)(4) did not contribute to or cause a serious injury or death of a patient or user.Pentax medical (b)(4) performed good faith effort to the user and confirmed the endoscope was used during the procedure, and there was no delay in the procedure which medical intervention was required.There was resistance felt during reprocessing and then the cleaning brush became stuck in the channel.On 30-apr-2021, pentax medical (b)(4) responded to an email request for additional information.They responded stating that the pentax medical cleaning brush model cs-6021t, unknown lot number, was damaged and clogged during reprocessing.Instructions for use(ifu), includes the following warning section 2-1-3, 3) "after using operational/cleaning accessories (e.G., forceps, needles, snares, brushes etc.) with the endoscope, carefully check that all accessories are intact and that no parts have fallen off and become lodged within the endoscope's instrument/suction channel.Furthermore, ensure that any therapeutic devices (e.G., clips, stents, etc.) passed through the channel are accounted for after use.On 06-apr-2016, pentax issued a u.S.Urgent field correction which is an ifu addendum for endoscopes with instrument channels.This addendum covers any operational/cleaning accessories and therapeutic devices which can become lodged in the endoscope's instrument channel.It reminds customers to carefully check that all accessories are intact, that no parts have fallen off and become lodged within the endoscope's instrument/suction channel and to ensure that any therapeutic devices (e.G., clips, stents, balloons, etc.) passed through the instrument channel and are accounted for after use.On 9-mar-2021, a device history record(dhr) review for model eg16-k10, serial number (b)(4)was performed under (b)(4), the dhr review confirmed the endoscope was manufactured on 18-sep-2012 under normal conditions, passed all required inspections, and was released accordingly.Also, there were no reworks or concessions and the dates of approval for shipment and actual date shipped were confirmed for 18-sep-2012.A device history record(dhr) review could not be completed for for pentax medical cleaning brush model cs-6021t as the lot number is unknown.During evaluation of the returned endoscope, a cleaning brush (cs6021t) was found stacked at the entrance of the forceps opening.The instructions for use(ifu) states that the forceps opening should be cleaned using cs-c9s, but this occurred because cs6021t was inserted.The endoscope was repair and approved on 19-may-2021.
 
Manufacturer Narrative
(b)(4).Mdr 9610877-2021-00102 will be submitted for the pentax medical video gastroscope model eg16-k10, serial number (b)(4).Mdr 9610877-2021-00103 will be submitted for the pentax medical cleaning brush model cs-6021t, unknown lot number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
VIDEO GASTROSCOPE
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0012
JA 
MDR Report Key12096345
MDR Text Key281610354
Report Number9610877-2021-00102
Device Sequence Number1
Product Code FDS
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial
Report Date 02/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEG16-K10
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 02/25/2021
Initial Date FDA Received06/30/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-