CAREFUSION SD 8300 ALARIS ETCO2 MODULE; PUMP, INFUSION Back to Search Results
Model Number 8300
Device Problems Circuit Failure (1089); Crack (1135); Communication or Transmission Problem (2896); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
This device was evaluated and repaired through the service repair process.Upon visual inspection the service technician noted that error code 570.6500 is confirmed due to a failed oridion board, replaced it a new oridion board.Updated a new logic board per c.O #(b)(4).Display board failed with error code 511.2010, replaced it a new display board.Damaged front case and rear case, replaced them new front and rear cases assembly.Performed leak down test and co2 calibration with passing results.Note: thank you for your patience for the part shortage.The failure code othe was used to track the alaris pump software version as received from the customer and the software version when device was sent back to the customer.It does not reflect a device failure or represent any risk to the patient.A review of the device history record showed the device had a manufacture date of 18jul2011.The review was performed from the date of manufacture to the date of product release for distribution.A review of the device history record in sap for sn (b)(4) was performed which confirmed that this device was not involved in a production failure which correlates to the customer reported issue.Based on the findings, service determined that the probable root cause of the reported issue was due to electrical failure of the oridion board communication failure.A review of the complaint history record in trackwise and sap was performed for the sn (b)(4) which did not confirm similar complaints with the same or related failure mode for this customer.
 
Event Description
The customer reported alarm - error codes / messages.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
8300 ALARIS ETCO2 MODULE
Type of Device
PUMP, INFUSION
Manufacturer (Section D)
CAREFUSION SD
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer (Section G)
CAREFUSION SD
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer Contact
brett wilko
10020 pacific mesa blvd
san diego, CA 92121-4386
8586172000
MDR Report Key12097846
MDR Text Key259385832
Report Number2016493-2021-509272
Device Sequence Number1
Product Code CCK
UDI-Device Identifier10885403830013
UDI-Public10885403830013
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031741
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 02/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/01/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8300
Device Catalogue Number8300
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/04/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/30/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/18/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-