ADVANCED BIONICS, LLC HIRES ULTRA IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1600-04
Device Problems Electrical /Electronic Property Problem (1198); Mechanical Problem (1384); Impedance Problem (2950)
Patient Problem Insufficient Information (4580)
Event Date 06/07/2021
Event Type  malfunction  
Event Description
The recipient reportedly experienced decreased performance.A review of the recipient¿s test data indicated impedance issues and open circuits.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed cuts in the silastic overmold on the top cover.In addition, the electrode was severed prior to receipt.This is believed to have occurred during revision surgery.The photographic imaging inspection revealed an electrode was broken within the electrode pocket.System lock was verified.The device passed some of the electrical tests performed.The device passed the mechanical test performed.The failure of this device is attributed to an electrode short in the electrode pocket.A corrective action was implemented.This version of the ultra device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES ULTRA IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
MDR Report Key12104826
MDR Text Key262031320
Report Number3006556115-2021-00863
Device Sequence Number1
Product Code MCM
UDI-Device Identifier07630016841002
UDI-Public(01)07630016841002(11)170215(17)200229
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup,Followup
Report Date 06/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/29/2020
Device Model NumberCI-1600-04
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2021
Initial Date Manufacturer Received Not provided
Initial Date FDA Received07/01/2021
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received08/11/2021
09/23/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age4 YR
-
-