Model Number CI-1600-04 |
Device Problems
Electrical /Electronic Property Problem (1198); Mechanical Problem (1384); Impedance Problem (2950)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 06/07/2021 |
Event Type
malfunction
|
Event Description
|
The recipient reportedly experienced decreased performance.A review of the recipient¿s test data indicated impedance issues and open circuits.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed cuts in the silastic overmold on the top cover.In addition, the electrode was severed prior to receipt.This is believed to have occurred during revision surgery.The photographic imaging inspection revealed an electrode was broken within the electrode pocket.System lock was verified.The device passed some of the electrical tests performed.The device passed the mechanical test performed.The failure of this device is attributed to an electrode short in the electrode pocket.A corrective action was implemented.This version of the ultra device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|